California Instruments 1251RP Series Manual de usuario Pagina 10

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 12
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 9
www.california-instruments.com
ISO 9001:2000
An Company
CTS Series - Report Generation
MS Word Test Reports
Test reports for harmonics, icker and
immunity tests are generated using MS
Word format. This widely used report
format can be integrated into more
elaborate user specic reports, in Word,
WordPerfect,orOpenOcecoveringall
aspects of compliance testing.
Test reports contain data on the EUT, the
test lab and operator, all measurement
results and a clear pass or fail indication.
Harmonics test reports include current
harmonics and voltage harmonics data in
both bar charts and tabular formats.
Detailed measurement data is also avail-
able on disk and can be exported to a
tab delimited ASCII text format for use
in other application programs such as
MS Excel. This allows further analysis
of the acquired data for engineering
troubleshooting purposes of EUTs that
did not pass.
Note: A copy of MS Word must be installed
on the PC to generate test reports.
The Power Analyzer and Conditioning System unit provides the required interface between the AC source, the Equipment Under
Test and the PC. A high current (75 Arms) version, PACS-3-75, is supplied with MX-CTS systems.
PACS Model: PACS-1 PACS-3 (PACS-3-75)
Number of phases 1 3
Channels Voltage and Current 2 6
Connector Style Front panel CEE/77 front none
Rear panel terminal block terminal block
Maximum voltage Front panel 240 Vac n/a
Rear panel 300 Vac 480 Vac
Maximum current Front panel 16 Arms n/a
Rear panel 40 Arms 40 Arms/ph (75 Arms/ph)
IEC 60725 Reference RP-Series CTS: internal n/a
Impedance
1
iX-Series CTS: programmable Z programmable Z
Optional Lumped: option -LR1 option -LR3/-LR4
Input Power Voltage 115 / 230 V ± 10 % 115 / 230 V ± 10 %
Current / Frequency < 0.5 A / 50 or 60 Hz < 0.75 A / 50 or 60 Hz
Dimensions HxWxD 3.5 x 16.8 x 22 3.5 x 16.8 x 22
HxWxD 89 x 427 x 560 89 x 427 x 560
1 MX-CTS systems do not oer programmable impedance and require option -LR4/OMNI-3-75 for Flicker test. Options -LR1 and -LR3
may be added to 3001iX-CTS / 5001iX-CTS or 15003iX-CTS congurations to be used in lieu of the standard programmable impedance.
Option -LR1 is built-in to PACS-1. Option -LR3 and -LR4 consist of OMNI-3-18i and OMNI-3-37i respectively. See OMNI data under op-
tions on last page.
CTS Series - PACS Specications
Vista de pagina 9
1 2 ... 5 6 7 8 9 10 11 12

Comentarios a estos manuales

Sin comentarios