CTS Series - Report Generation
MS Word Test Reports
Test reports for harmonics, flicker and
immunity tests are generated using MS
Word format. This widely used format
report can be integrated into more
elaborate reports covering all aspects
of compliance testing if needed.
Test reports contain data on the EUT,
the test lab and operator, all measure-
ment results and a clear pass or fail
indication. Harmonics test reports in-
clude current harmonics and voltage
harmonics data in both bar charts and
tabular formats.
Detailed measurement data is also
available on disk and can be exported
to a tab delimited ASCII text format for
use in other application programs such
as MS Excel. This allows further analy-
sis of the acquired data for engineer-
ing troubleshooting purposes of EUT’s
that did not pass.
Note: A copy of MS Word must be in-
stalled on the PC to generate test re-
ports.
The Power Analyzer and Conditioning System unit provides the required interface between the AC source, the Equipment
Under Test and the PC. Single phase CTS Systems are supplied with a PACS-1, three phase CTS systems are supplied
with a PACS-3. RP Series based CTS systems include a lumped reference impedance in the PACS-1
PACS Model: PACS-1 PACS-3
Number of phases 1 3
Channels Voltage and Current 2 6
Connector Style Front panel CEE/77 front none
Rear panel terminal block terminal block
Maximum voltage Front panel 240 Vac n/a
Rear panel 300 Vac 480 Vac
Maximum current Front panel 16 Arms n/a
Rear panel 40 Arms 40 Arms/phase
IEC 725 Reference RP-Series CTS: internal n/a
Impedance
1
iX-Series CTS: programmable Z programmable Z
Optional Lumped: option -LR1 option -LR3/-LR4
Dimensions HxWxD 3.5 x 16.8 x 22 5.25 x 16.8 x 22
HxWxD 89 x 427 x 560 134 x 427 x 560
1 30003iX-CTS systems do not offer programmable impedance and require option -LR4 for Flicker test. Options -LR1
and -LR3 may be added to 3001iX-CTS / 5001iX-CTS or 15003iX-CTS configurations to be used in lieu of the
standard programmable impedance. Option -LR1 is built-in to PACS-1. Option -LR3 and -LR4 consist of OMNI-3-18i
and OMNI-3-37i respectively. See OMNI data in options section of this catalog.
CTS Series - PACS Specifications