Output Current
Wire Size
Maximum wire length from output to load, 1% drop
5.0 A
RMS
0.75 mm
2
6 m 20 ft
5.0 A
RMS
AWG 18
6.4 m 21 ft
5.0 A
RMS
1.0 mm
2
8 m 26 ft
5.0 A
RMS
AWG 16
8.5 m 28 ft
5.0 A
RMS
1.5 mm
2
12 m 39 ft
5.0 A
RMS
AWG 14
14 m 45 ft
Table 3-1: Maximum Output Wiring Lengths
The Remote Sense inputs must be connected or an output voltage 7% higher than the
programmed output will be generated. If the 2003RP is configured for constant voltage, a
fault will be generated.
L
N
GND
Ferrite
L N Gnd
AC Input
Safety Cover
AC
Input
HI LO
AC
Output
Ferrite
Clamp
AC Output
Safety Cover
Figure 3-6: Input and Output Wiring