California Instruments 801RP Series Especificaciones Pagina 91

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 126
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 90
CALibrate:MEASure:POWer[:FSCale] <NRf>
This command affects the calibration of the power measurement at full scale.
Command Syntax CALibrate:MEASure:POWer[:FSCale] <NRf>
Parameters <NRf> (value range -127 to +127)
Examples CAL:MEAS:POW 78
Query Syntax CALibrate:MEASure:POWer?
Returned Parameters <NR1> (value range -127 to +127)
Related Commands CALibrate:MEASure:CURRent
CALibrate:MEASure:POWer:ZERO <NRf>
This command affects the calibration of the power measurement zero offset.
Command Syntax CALibrate:MEASure:POWer:ZERO <NRf>
Parameters <NRf> (value range -127 to +127)
Examples CAL:MEAS:POW:ZERO -5
Query Syntax CALibrate:MEASure:POWer:ZERO?
Returned Parameters <NR1> (value range -127 to +127)
Related Commands CALibrate:MEASure:POWer
CALibrate:MEASure:VOLTage[:FSCale] <NRf>
This command affects the calibration of the rms voltage measurement at full scale.
Command Syntax CALibrate:MEASure:VOLTage[:FSCale] <NRf>
Parameters <NRf> (value range -127 to +127)
Examples CAL:MEAS:VOLT 56
Query Syntax CALibrate:MEASure:VOLTage?
Returned Parameters <NR1> (value range -127 to +127)
Related Commands CALibrate:MEASure:CURRent
CALibrate:MEASure:VOLTage:ZERO <NRf>
This command affects the calibration of the rms voltage measurement zero offset.
Command Syntax CALibrate:MEASure:VOLTage:ZERO <NRf>
Parameters <NRf> (value range -127 to +127)
Examples CAL:MEAS:VOLT:ZERO +4
Query Syntax CALibrate:MEASure:VOLTage:ZERO?
Returned Parameters <NR1> (value range -127 to +127)
Related Commands CALibrate:MEASure:VOLTage
Vista de pagina 90