MX / RS Series SCPI Programming Manual AMETEK Programmable Power
MX / RS Series 123
4.26 Trigger Subsystem
This subsystem controls the triggering of the AC/DC power source. See chapter 6 under
"Triggering Output Changes" for an explanation of the Trigger Subsystem. The INITiate
commands control the initialization of the transient system. The trigger subsystem must first
be enabled using the INITiate commands or no triggering action will occur. Refer to Figure
6-5 for a block diagram of the trigger system.
Subsy
stem Syntax
ABORt Resets the trigger system to the Idle state
INITiate
[:IMMediate] Initiates the system for one trigger
[:TRANsient]
:ACQuire
:CONTinuous sets if the initate continuous to (ON|OFF)
TRIGger
[:TRANsient]
:SOURce Sets the trigger source (IMMediate|BUS|EXT)
:COUNt Set if the trigger is controlled at every
count(NONE|ALL)
:ACQuire
:SOURce set the aquire trigger soure
(IMMediate|SYNChronize|BUS|TTLTrg)
:SYNChronize
:SOURce <source> Sets the trigger sync source (IMM|PHAS)
:PHASe <n> Sets the synchronous phase reference
:STATe? Return the trigger state (IDLE|ARM|BUSY)
ABORt
This command resets the transient trigger systems to the Idle state. Any output transient or
measurement that is in progress is immediately aborted. ABORt also cancels any lists or
pulses that may be in process.
Command Syntax ABORt
Parameters one
Examples ABOR
Related Commands INIT *RST *TRG