CALibrate:MEASure:CURRent:TEMPerature:DC[:FSCale] <NRf>
This command initiates the calibration of the DC current measurement at elevated
temperature.
Command Syntax CALibrate:MEASure:CURRent:TEMPerature:DC[:FSCale]
Parameters <NRf> (0 or desired offset value)
Examples CAL:MEAS:CURR:TEMP:DC
Query Syntax CALibrate:MEASure:CURRent:TEMPerature:DC?
Returned Parameters <NR1> (value range -1000 to + 1000
Related Commands CALibrate:MEASure:CURRent[:AMB]:DC[:FSCale]
CALibrate:MEASure:CURRent:TEMPerature:DC:ZERO <NRf>
This command initiates the offset adjustment of the DC current measurement at elevated
temperature.
Command Syntax CALibrate:MEASure:CURRent:TEMPerature:DC:ZERO
Parameters <NRf> (0 or desired offset value)
Examples CAL:MEAS:CURR:TEMP:DC:ZERO
Query Syntax CALibrate:MEASure:CURRent:TEMPerature:DC:ZERO?
Returned Parameters <NR1> (value range 0 to +5)
Related Commands CALibrate:MEASure:CURRent[:AMB]:DC:ZERO