AMETEK Programmable Power MX / RS Series SCPI Programming Manual
44 MX / RS Series
CALibrate:MEASure:VOLTage:TEMPerature:DC[:FSCale] <NRf>
This command initiates the calibration of the DC voltage measurement at full scale and at
elevated temperature.
Command Syntax CALibrate:MEASure:VOLTage:TEMPerature:DC[:FSCale]
Parameters <NRf> (actual DC output voltage measured with external device)
Examples CAL:MEAS:VOLT:TEMP 120
Query Syntax CALibrate:MEASure:VOLTage:TEMPerature:DC?
Returned Parameters <NR2> (value range -1000 to +1000)
Related Commands CALibrate:MEASure:VOLTage[:AMBient]:DC[:FSCale]
CALibrate:MEASure:VOLTage:TEMPerature:DC:ZERO <NRf>
This command initiates the offset adjustment of the rms voltage measurement at elevated
temperature.
Command Syntax CALibrate:MEASure:VOLTage:TEMPerature:DC:ZERO
Parameters <NRf> (0 or desired offset value)
Examples CAL:MEAS:VOLT:TEMP:DC:ZERO 0
Query Syntax CALibrate:MEASure:VOLT:DC:ZERO?
Returned Parameters <NR2> (value range -20 to +20)
Related Commands CALibrate:MEASure:VOLTage[:AMBient]:DC:ZERO