Programming Manual Compact i / iX Series
AMETEK Programmable Power Division 112
TRIGger:SEQuence2:SOURce
TRIGger:SYNChronize:SOURce
These commands select the synchronizing trigger source in generating a step, pulse, or list
output as follows:
IMMediate starts the transient output immediately, unless a delay time other than 0
has been specified by TRIGger:DELay. In this case the transient output
starts after the expiration of the delay time.
PHASe starts the transient output at the reference phase set by
TRIG:SYNC:PHAS.
Command Syntax TRIGger:SEQuence2:SOURce<CRD>
TRIGger:SYNChronize:SOURce<CRD>
Parameters IMMediate | PHASe
*RST Value IMM
Examples TRIG:SYNC:SOUR IMM TRIG:SEQ2:SOUR PHAS
Query Syntax TRIGger:SEQuence2:SOURce?
TRIGger:SYNChronize:SOURce?
Returned Parameters <CRD>
Related Commands ABOR TRIG:DEL TRIG:SYNC
TRIG:SYNC:PHAS INIT INIT:CONT *TRG *WAI
TRIGger:SEQuence2:PHASe
TRIGger:SYNCHronize:PHASe
These commands set the phase angle with respect to an internal phase reference at which
PHASe:SYNChronous:SOURce becomes true. The range is from -360 to +360 degrees.
Command Syntax TRIGger:SEQuence2:PHASe<NRf+>
TRIGger:SYNChronize:PHASe<NRf+>
Parameters -360° to +360°
*RST Value 0
Examples TRIG:SYNC:PHAS 90 TRIG:SEQ2:PHAS 180
Query Syntax TRIGger:SEQuence2:PHASe?
TRIGger:SYNChronize:PHASe?
Returned Parameters <NR3>
Related Commands ABOR TRIG:DEL TRIG:SYNC
TRIG:SYNC:PHAS INIT INIT:CONT *TRG *WAI
TRIGger:SEQuence3
TRIGger:ACQuire
When the trigger subsystem has been initiated, these commands generate a measurement
trigger regardless of the selected trigger source. The measurement trigger causes the AC source
to digitize the instantaneous output voltage and current for several output cycles and store the
results in a buffer.
The FETCh commands return the requested calculation from this acquired data. When the
measurement completes, the WTG bit in the Status Operation Condition register is cleared.