Programming Manual Compact i / iX Series
AMETEK Programmable Power Division 211
Number Message String Cause Remedy
815 "Slew time exceed dwell" Time needed to slew
to final value is less
than dwell time.
Check dwell times in transient list
settings. Increase dwell time or change
slew rate for affected parameter.
816 "Illegal during transient" Operation requested
not available while
transient is running.
Wait till transient execution is completed
or abort transient execution first.
817 "Output relay must be
closed"
Transient
programmed with
output relay open.
Close relay before attempting transient
operation.
819 "Clock and sync must be
internal"
Operation not
possible with external
clock
Switch to internal sync. (Default)
820 "Input buffer full" Too much data
received.
Break up data in smaller blocks.
821 "PFC Input Fault " AC input line related
error.
Possibly due to low line input condition
requiring too much input current.
822 "Waveform harmonics
limit"
Harmonic contents of
user defined wave
shape is too high and
could damage
amplifier output
stage.
Reduce harmonic content or reduce
fundamental frequency programmed.
823 "Amplifier fault" An amplifier failure.
Can be reported at
any time.
Determine which amplifier is at fault with
self-test or checking LED on Relay
Board. Replace amplifier.
824 “Output relay must be
open”
Operation requested
not possible with
relay state closed.
Generaly caused by trying to change
voltage range while output relay is
closed. Open relay first, then voltage
range can be changed.
825 “Over voltage prot trip” Over voltage
detected on output
Check output voltage for correct RMS
value.
826 “Peak current prot trip” Peak current limit
exceeded.
Peak current exceeded. Could be
caused by switching EUT on or off.
827 “Frequency error” Frequency error
during self-test.
Correct frequency was not measured
during self-test. May be result of 801
error.
828 “Phase error” Self test error phase
angle
Correct phase angle was not measured
during self-test. May be result of 801
error.
829 “Dc component exceed
limit”
Too much DC content
in loaded ARB
waveform.
Check waveform programming.
Table 8-3: SCPI error codes and messages.