Programming Manual Compact i / iX Series
AMETEK Programmable Power Division 56
Query Syntax MEASure[:SCALar]:POWer:AC:APParent?
FETCh[:SCALar]:POWer:AC:APParent?
Parameters None
Examples MEAS:POW:AC:APP? FETC:POW:AC:APP?
Returned Parameters <NR3>
Related Commands INST:NSEL
MEASure:POWer:AC:REACtive?
FETCh:POWer:AC:REACtive?
Phase Selectable
These queries return the reactive power being sourced at the output terminals in volt-amperes
reactive. Reactive power is computed as:
VAR = sqrt(square(apparent power) - square(real power))
Query Syntax MEASure[:SCALar]:POWer:AC:REACtive?
FETCh[:SCALar]:POWer:AC:REACtive?
Parameters None
Examples MEAS:POW:AC:REAC? FETC:POW:AC:REAC?
Returned Parameters <NR3>
Related Commands INST:NSEL
MEASure:POWer:AC:PFACtor?
FETCh:POWer:AC:PFACtor?
Phase Selectable
These queries return the output power factor. The power factor is computed as:
pfactor = real power/apparent power
Query Syntax MEASure[:SCALar]:POWer:AC:PFACtor?
FETCh[:SCALar]:POWer:AC:PFACtor?
Parameters None
Examples MEAS:POW:AC:PFAC? FETC:POW:AC:PFAC?
Returned Parameters <NR3>
Related Commands INST:NSEL
MEASure:POWer:AC:TOTal?
FETCh:POWer:AC:TOTal?
These queries return the total power being sourced at the output terminals of a three-phase AC
source. Not applicable on single phase models or 3 phase model with –MODE option when in
single phase mode.
Query Syntax MEASure[:SCALar]:POWer:AC:TOTal?
FETCh[:SCALar]:POWer:AC:TOTal?
Parameters None
Examples MEAS:POW:AC:TOT? FETC:POW:AC:TOT?
Returned Parameters <NR3>
Vista de pagina 55