Programming Manual Compact i / iX Series
AMETEK Programmable Power Division 209
Number Message String Cause Remedy
-255 “Directory full” Too many waveform
directory entries
Delete one or more waveforms from
waveform memory to make room.
-256 “File name not found” Waveform requested
not in directory
Check waveform directory for waveform
names present.
-257 “File name error” Incorrect filename Too many or non ASCII characters used
in waveform file definition.
-283 “Illegal variable name” Variable name illegal. Use ASCII characters only
-300 "Device specific error" Hardware related
error
Check hardware for proper operation.
-311 "Memory error" Waveform memory
checksum error.
May be the result of incomplete user-
defined waveform download. Check
interface and try downloading waveform
again. Successful download may clear
this error condition.
Alternatively, use TRAC:DEL:ALL
command to clear waveform memory.
-314 "Save/recall memory
lost"
User setup register
contents lost
Store setup in same register again.
-315 "Configuration memory
lost"
Hardware
configuration settings
lost.
Contact CI service department at
to obtain
instructions on restoring configuration
data.
-330 "Self-test failed" Internal error Contact CI service department at
-350 "Queue overflow" Message queue full. Too many message. Read status using
SYST:ERR query until 0, "No Error" is
received indicating queue empty.
-400 "Query error" Unable to complete
query.
Check programming manual for correct
query format and parameters
-410 "Query INTERRUPTED" Query issued but
response not read.
Check application program for correct
flow. Response must be read after each
query to avoid this error.
-420 "Query
UNTERMINATED"
Query incomplete. Check for terminator after query
command.
-430 "Query DEADLOCKED" Query cannot be
completed
Check application program for multiple
queries
-440 "Query
UNTERMINATED"
Query incomplete. Check for terminator after query
command.
0 "No error" No errors in queue
40 "Voltage self test error,
output 1”
No. 1/A amplifier in
Master source has no
or incorrect output
voltage during Self-
test.
Contact CI service department at
41 "Voltage self test error,
output 2”
No. 2/B amplifier in
Master source has no
or incorrect output
voltage during Self-
test. Three phase
models only.
Contact CI service department at
42 "Voltage self test error, No. 3/C amplifier in Contact CI service department at