
Programming Manual Compact i / iX Series
AMETEK Programmable Power Division 165
DO160:NORMal:VOLTage:INTerrupt
This command will cause output voltage interruption. There are 15 different levels of interruption.
Refer to Figure 8-6 in the User Manual for characteristics of each level. Test numbers 16 and 17
for all equipment that does not incorporate digital circuit. Test numbers 21 to 26 correspond to
Standard I through VI for EUROCAE and RTCA2. See Figure 8-7 in User manual.
Command Syntax DO160:NORMal:VOLTage:INTerrupt <NRf>
Parameters <test number>
Examples DO160:NORM:VOLT:INT 4
Query Syntax DO160:NORM:VOLT:INT?
Returned Parameters <NR1>
DO160:NORMal:VOLTage:SURGe
This command will generate the voltage levels required to generate a normal voltage source.
Refer to Section 8.1.4.1, under the heading titled
VOLTAGE SURGE, for details.
Command Syntax DO160:NORMal:VOLTage:SURGe
Parameters none
Examples DO160:NORM:VOLT:SURG
Query Syntax none
DO160:NORMal:FREQuency:MODulation
This command will cause output frequency modulation. The level of modulation is the function of
the rate of modulation. Refer to Figure 8-5 in the User Manual. The command parameter is the
rate of modulation in Hz.
Command Syntax DO160:NORMal:FREQuency:MODulation <NRf>
Parameters <frequency>
Examples DO160:NORM:FREQ:MOD 0.1
Query Syntax DO160:NORM:FREQ:MOD?
Returned Parameters <NR2>
DO160:NORMal:FREQuency:TRANsient
This command applies to Group 1 only for the the Eurocae standard. Refer to Table 8-5 in the
User Manual.
Command Syntax DO160:NORMal:FREQuency:TRANsient
Parameters none
Examples DO160:NORM:FREQ:TRAN
Query Syntax none
Returned Parameters none
Comentarios a estos manuales