Programming Manual Compact i / iX Series
AMETEK Programmable Power Division 208
Number Message String Cause Remedy
-120 "Numeric data error" Data received is not a
number
Check programming manual for correct
command syntax
-121 "Invalid character in
number"
Number received
contains non-numeric
character(s)
Check programming manual for correct
command syntax
-123 "Exponent too large" Exponent in number
exceeds limits
Check programming manual for correct
parameter range
-128 "Numeric data not
allowed"
Number received
when number is not
allowed.
Check programming manual for correct
command syntax
-168 "Block data not allowed" Block data was sent. Check programming manual for correct
command syntax
-200 "Execution error" Command could not
be executed
Command may be inconsistent with
mode of operation.
-201 "Invalid while in local" Command issued but
unit is not in remote
state
Put instrument in remote state before
issuing GPIB commands.
-203 "Command protected" Command is locked
out
Some commands are supported by the
unit but are locked out for protection of
settings and are not user accessible.
-210 "Trigger error" Problem with trigger
system.
Unit could not generate trigger for
transient execution or measurement.
-211 "Trigger ignored" Trigger request has
been ignored.
Trigger setup incorrect or unit was not
armed when trigger was received.
Check transient system or
measurement trigger system settings.
-213 "Init ignored" Initialization request
has been ignored
Unit was told to go to armed state but
was unable to do so. Could be caused
by incorrect transient system or
measurement acquisition setup.
-220 "Parameter error" Parameter not
allowed.
Incorrect parameter or parameter value.
Check programming manual for
allowable parameters
-221 "Setting conflict" Transient
programmed with
more than 1 mode.
Selftest setting
conflict
Check other settings. E.g. Redefine
transient mode.
As result of *TST? execution, indicates
ALC mode is off or waveform not set to
Sine.
-222 "Data out of range" Parameter data
outside of allowable
range.
Check programming manual for
allowable parameter values
-223 "Too much data" More data received
than expected
Check programming manual for number
of parameters or data block size
-224 "Illegal parameter value" Parameter value is
not suppored
Check programming manual for correct
parameters
-226 "Lists not same length" One or more transient
lists programmed has
different length.
All lists must be of same length or
transient cannot be compiled and
executed.
-241 "Hardware missing" N/A N/A
-254 "Media full" No storage space left
to save settings or
data.
Delete other settings or data to make
room.