Programming Manual Compact i / iX Series
AMETEK Programmable Power Division 156
7.6 Status Byte Register
This register summarizes the information from all other status groups as defined in the IEEE
488.2 Standard Digital Interface for Programmable Instrumentation. The bit configuration is
shown in Table 5-3.
Command Action
*STB? reads the data in the register but does not clear it (returns MSS in bit 6)
serial poll reads and clears the data in the register (returns RQS in bit 6)
The MSS Bit
This is a real-time (unlatched) summary of all Status Byte register bits that are enabled by the
Service Request Enable register. MSS is set whenever the AC source has one or more reasons
for requesting service. *STB? reads the MSS in bit position 6 of the response but does not clear
any of the bits in the Status Byte register.
The RQS Bit
The RQS bit is a latched version of the MSS bit. Whenever the AC source requests service, it
sets the SRQ interrupt line true and latches RQS into bit 6 of the Status Byte register. When the
controller does a serial poll, RQS is cleared inside the register and returned in bit position 6 of the
response. The remaining bits of the Status Byte register are not disturbed.
The MAV bit and Output Queue
The Output Queue is a first-in, first-out (FIFO) data register that stores AC source-to-controller
messages until the controller reads them. Whenever the queue holds one or more bytes, it sets
the MAV bit (bit 4) of the Status byte register.