Programming Manual Compact i / iX Series
AMETEK Programmable Power Division 152
Bit Signal Meaning
Operation Status Group
0 CAL Interface is computing new cal constants
5 WTG Interface is waiting for a trigger
8 CV The output voltage is regulated
Questionable and Questionable Instrument Isummary Status
Groups
0 OV The overvoltage protection circuit has tripped
1 OCP The overcurrent protection circuit has tripped
3 UNR The output is unregulated
4 TO An overtemperature condition has occurred
9 RI The remote inhibit state is active
11 Rail Loss of input phase voltage detected
12 CLrms The rms current limit circuit is active
13 Isum Summary of QUES:INST:ISUM registers
Standard Event Status Group
0 OPC Operation complete
2 QYE Query error
3 DDE Device-dependent error
4 EXE Execution error
5 CME Command error
7 PON Power-on
Status Byte and Service Request Enable Registers
3 QUES Questionable status summary bit
4 MAV Message Available summary bit
5 ESB Event Status Summary bit
6 MSS Master Status Summary bit
RQS Request Service bit
7 OPER Operation status summary bit
Table 7-2: Bit Configurations of Status Registers