5.7 *RCL
Warning Recalling a previously stored state may place hazardous voltages at the AC
source output.
This command restores the AC source to a state that was previously stored in memory with a
*SAV command to the specified location. All states are recalled with the following exceptions:
• CAL:STATe is set to OFF
• the trigger system is set to the Idle state by an implied ABORt command (this cancels any
uncompleted trigger actions)
The device state stored in location 0 is automatically recalled at power turn-on when the
OUTPut:PON:STATE is set to RCL0. This register only saves steady state setting, not the
transient list data. All other registers (1-15) saved both the steady state settings and the transient
list data record.
Command Syntax *RCL <NRf>
Parameters 0 through 15
Example *RCL 3
Related Commands *PSC *RST *SAV
Note that recalling settings and transient list information from non-volatile memory takes some
amount of time as shown in the table below. If you develop a test program that uses the *RCL or
*SAV commands, insert the amount of delay shown in the table before sending the next
command in your program.
Command: *RCL 0 *RCL n (n = 1 to 15) *SAV 0 *SAV n (n = 1 to 15)
Execution time: 20 msec 40 msec 80 msec 40 msec