Programming Manual Compact i / iX Series
AMETEK Programmable Power Division 118
5.7 *RCL
Warning Recalling a previously stored state may place hazardous voltages at the AC
source output.
This command restores the AC source to a state that was previously stored in memory with a
*SAV command to the specified location. All states are recalled with the following exceptions:
CAL:STATe is set to OFF
the trigger system is set to the Idle state by an implied ABORt command (this cancels any
uncompleted trigger actions)
The device state stored in location 0 is automatically recalled at power turn-on when the
OUTPut:PON:STATE is set to RCL0. This register only saves steady state setting, not the
transient list data. All other registers (1-15) saved both the steady state settings and the transient
list data record.
Command Syntax *RCL <NRf>
Parameters 0 through 15
Example *RCL 3
Related Commands *PSC *RST *SAV
Note that recalling settings and transient list information from non-volatile memory takes some
amount of time as shown in the table below. If you develop a test program that uses the *RCL or
*SAV commands, insert the amount of delay shown in the table before sending the next
command in your program.
Command: *RCL 0 *RCL n (n = 1 to 15) *SAV 0 *SAV n (n = 1 to 15)
Execution time: 20 msec 40 msec 80 msec 40 msec