Programming Manual Compact i / iX Series
AMETEK Programmable Power Division 164
8.2.2 DO160 Normal Test Commands
DO160:NORMal:VOLT_FREQ:MINimum
This command will set the voltage and frequency to the minimum level for the normal operation.
Command Syntax DO160:NORMal:VOLT_FREQ:MINimum
Parameters none
Examples DO160:NORM:VOLT_FREQ:MIN
Query Syntax none
DO160:NORMal:VOLT_FREQ:MAXimum
This command will set the voltage and frequency to the maximum level for the normal operation.
Command Syntax DO160:NORMal:VOLT_FREQ: MAXimum
Parameters none
Examples DO160:NORM:VOLT_FREQ:MAX
Query Syntax none
DO160:NORMal:VOLTage:UNBalance
This command will unbalance the voltage level for each phase to the normal high and low
operating voltage of each phase. This test is valid only for three phase AC source.
Command Syntax DO160:NORMal:VOLTage:UNBalance
Parameters none
Examples DO160:NORM:VOLT:UNB
Query Syntax none
DO160:NORMal:WAVeform:DISTortion
This command will set the voltage distortion to 5% for the duration of the test.
Command Syntax DO160:NORMal:WAVeform:DISTortion
Parameters none
Examples DO160:NORM:WAV:DIST
Query Syntax none
DO160:NORMal:VOLTage:MODulation
This command will cause output voltage modulation. The level of modulation is the function of the
rate of modulation. Refer to Figure 8-4 in the User Manual. The command parameter is the rate
of modulation in Hz.
Command Syntax DO160:NORMal:VOLTage:MODulation <NRf>
Parameters <frequency>
Examples DO160:NORM:VOLT:MOD 20
Query Syntax DO160:NORM:VOLT:MOD?
Returned Parameters <NR1>