California Instruments 2253iX Manual de instrucciones Pagina 55

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 216
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 54
Programming Manual Compact i / iX Series
AMETEK Programmable Power Division 55
4.11 Power Measurement Subsystem
This subsystem programs the power measurement capability of the power source. Two
measurement commands are available: MEASure and FETCh. MEASure triggers the acquisition
of new measurement data before returning a reading. FETCh returns a reading computed from
previously acquired data.
Individual outputs of a three-phase source are specified by the setting of INSTrument:NSELect.
Subsystem Syntax
MEASure | FETCh
[:SCALar]
:POWer
[:DC]? Returns the dc component of power
:AC
[:REAL]? Returns real power
:APParent? Returns VA
:REACtive? Returns VAR
:PFACtor? Returns power factor
:TOTal? Returns real 3-phase total power
MEASure:POWer?
FETCh:POWer?
Phase Selectable
These queries return the dc component of the power being sourced at the output terminals in
watts.
Query Syntax MEASure[:SCALar]:POWer[:DC]?
FETCh[:SCALar]:POWer[:DC]?
Parameters None
Examples MEAS:POW? FETC:POW?
Returned Parameters <NR3>
Related Commands INST:NSEL
MEASure:POWer:AC?
FETCh:POWer:AC?
Phase Selectable
These queries return the in-phase component of power being sourced at the output terminals in
watts.
Query Syntax MEASure[:SCALar]:POWer:AC[:REAL]?
FETCh[:SCALar]:POWer:AC[:REAL]?
Parameters None
Examples MEAS:POW:AC? FETC:POW:AC?
Returned Parameters <NR3>
Related Commands INST:NSEL
MEASure:POWer:AC:APParent?
FETCh:POWer:AC:APParent?
Phase Selectable
These queries return the apparent power being sourced at the output terminals in volt-amperes.