Programming Manual Compact i / iX Series
AMETEK Programmable Power Division 117
5.5 *OPT?
This query requests the AC Source to identify any options that are installed. Options are identified
by number. A 0 indicates no options are installed.
Query Syntax *OPT?
Returned Parameters <AARD>
5.6 *PSC
This command controls the automatic clearing at power-on of the Service Request Enable and
the Standard Event Status Enable registers (see chapter 7 under Programming the Status and
Event Registers for register details):
*PSC ON | 1
prevents the register contents from being saved causing them to be
cleared at power-on. This prevents a PON event from generating SRQ at
power-on.
PSC OFF | 0 saves the contents of the Standard Event Enable and Service Request
Enable registers in nonvolatile memory and recalls them at power-on. This
allows a PON event to generate SRQ at power-on.
Command Syntax *PSC<bool>
Parameters 0 | 1 | OFF | ON
Example *PSC 0 *PSC 1
Query Syntax *PSC?
Returned Parameters 0 | 1
Related Commands *ESE *SRE