Programming Manual Compact i / iX Series
AMETEK Programmable Power Division 90
VOLTage:PROTection
Phase Selectable
This command sets the overvoltage protection (OVP) level of the AC source. If the peak output
voltage exceeds the OVP level, then the AC source output is disabled and the Questionable
Condition status register OV bit is set (see Section 7 under Programming the Status and Event
Registers). An overvoltage condition can be cleared with the OUTPut:PROTection:CLEar
command after the condition that caused the OVP trip is removed. The OVP always trips with
zero delay and is unaffected by the OUTPut:PROTection:DELay command.
Command Syntax
[SOURce:]VOLTage:PROTection[:LEVel]<NRf+>
Parameters 0 to 500
Unit V (peak voltage)
*RST Value MAX
Examples VOLT:PROT 400 VOLT:PROT:LEV MAX
Query Syntax [SOURce:]VOLTage:PROTection[:LEVel]?
Returned Parameters <NR3>
Related Commands OUTP:PROT:CLE
VOLTage:RANGe
Phase Selectable
This command sets the voltage range of the AC source. Two voltage ranges are available: a 150
volt range and a 300 volt range. Sending a parameter greater than 150 selects the 300 volt
range, otherwise the 150 volt range is selected.
When the range is set to 150, the maximum rms voltage that can be programmed for a sine wave
is 150 volts. For other waveshapes, the maximum programmable voltage may be different,
depending on the waveform crest factor.
Note: The VOLTage:RANGe command is coupled with the CURRent command. This means
that the maximum current limit that can be programmed at a given time depends on the voltage
range setting in which the unit is presently operating.
Note: The VOLTage:RANGe command is also coupled with the OUTPUT command. A voltage
range change can only be accomplished if the output relay is open. If the output relay is closed,
sending a VOLT:RANG <n> command will generate an 824 Error and the voltage range will not
change. (Firmware 0.26 and higher only.)
Refer to chapter 4 under "Coupled Commands" for more information.
A user settable delay may be inserted between dropping the output voltage to zero and changing
the range relay state. See the “PONSetup:RELay” command for details. Your application
program should allow for this delay. (default is 0.1 sec or 100 msec).
Note: Change
Command Syntax [SOURce:]VOLTage:RANGe<NRf+>
Parameters 150 | 300
*RST Value MAX
Examples VOLT:RANG 150 VOLT:RANG MIN
Query Syntax [SOURce:]VOLTage:RANGe?
Returned Parameters <NR3>
Related Commands VOLT OUTP PONS:REL